A Gage Study Percentage Needs Its Denominator

Compare gauge R&R contribution, study variation and percent tolerance by checking their denominators and measured feature.

Two suppliers both report “10% gage R&R” for a shaft measurement. One divided measurement study variation by the variation among the observed measurements. The other divided by the customer's tolerance width. Both percentages can appear in one analysis, and the buyer needs to know which was used before comparing suppliers or accepting either inspection process.

Ask for the characteristic, method, part and operator design, variance components, calculation rule and denominator. Retain the total tolerance from the approved drawing and an explicit definition of which production condition the study represents. A report with only a traffic-light verdict leaves the most important comparison unauditable.

Identify what the study measures

Repeatability describes variation when the same operator repeats a measurement of the same part under the agreed setup. Reproducibility reflects additional measurement variation involving people or other conditions in the study design. A crossed study normally has each operator measure each study part. That structure allows an analysis to estimate differences between parts and sources of measurement variation. The details can include a part-by-operator interaction depending on the agreed method and observed design.

In a destructive or irreversible measurement, different operators may be unable to measure the same physical part. Decide on a design that respects that constraint and record what was treated as comparable. A crossed analysis that silently treats different parts as identical can assign real part variation to the measurement process. NIST’s gauge R&R chapter discusses artifacts, operators, configurations, repeatability, reproducibility and other measurement characteristics.

Gage R&R covers a defined measurement system and environment. Tool calibration status, bias over the measurement range and long-term stability supply related evidence. Ask for a second source of evidence when the decision depends on one of those properties. A gauge may repeat tightly while consistently reading high, or perform well in a laboratory yet have a different fixture at the station to be qualified.

Interpret three percentages in the same report

Minitab’s crossed gage R&R interpretation distinguishes variance contribution, percentage of study variation and percentage of tolerance. Before transferring a number to a supplier scorecard, write the numerator and denominator in the margin. Variance has squared units and standard deviation has the original units. Dividing the corresponding quantities produces different percentages.

Report fieldCalculation to identifyWhat it answers
Percent contributionMeasurement variance divided by total varianceHow much of modelled variance belongs to measurement
Percent study variationMeasurement study spread divided by total study spreadHow large measurement spread is compared with the spread represented by the study
Percent toleranceMeasurement study spread divided by tolerance widthHow much of the agreed specification width the measurement spread occupies

Check the multiplier used for study spread, especially when comparing reports prepared with different methods. If a study multiplies both standard deviations by the same constant, that multiplier cancels in the ratio for percent study variation. It still affects percent tolerance. The drawing tolerance must be expressed in the same units and as the agreed width. For a two-sided specification from 11.90 to 12.10 mm, the width is 0.20 mm.

Work through two fictional studies

For this exercise, repeatability standard deviation is 0.008 mm and reproducibility standard deviation is 0.006 mm, with no other measurement variance term in the fictional model. Combine variance components before taking the root. The total measurement standard deviation is `sqrt(0.008² + 0.006²)` or 0.010 mm. At six times standard deviation, the measurement study spread is 0.060 mm. Dividing it by the agreed width of 0.20 mm gives 30.0% of tolerance.

Now hold the fictional measurement performance fixed while changing the parts used in the study. Study A has a part-to-part standard deviation of 0.020 mm. Study B has 0.050 mm. Assume the stated variance components add in each simple model.

Fictional resultStudy AStudy B
Measurement standard deviation0.010 mm0.010 mm
Part-to-part standard deviation0.020 mm0.050 mm
Measurement contribution to total variance20.0%3.85%
Measurement share of study variation44.7%19.6%
Measurement share of the 0.20 mm tolerance30.0%30.0%

In Study A the total standard deviation is about 0.02236 mm. Study B gives about 0.05099 mm. These different denominators explain the two study-variation results with the same measurement performance and tolerance. Changing the sample of parts can change a relative metric. Keep the sampling rationale, actual part IDs and the units before drawing a conclusion about the gauge. These figures are examples, and acceptance thresholds must come from the parties' agreed criteria.

Bind the study to the actual purchasing decision

Start the order with the measured feature and drawing revision. Require the inspection method and gauges intended for the actual line, a proposed study design, the calculation and a clear criterion. Ask who addresses a discrepancy if the current fixture, part geometry or line changes. If a report was accepted on a legacy part family, look for evidence that the new feature can reasonably use the same measurement system and conditions.

In a supplier demonstration, request one coded part record and find all its readings across operators and repeated measurements. Ask how readings were ordered and whether the operator could recognize previous results. Then choose a reported percentage and recompute it from the components. A difference can come from a different tolerance revision, an omitted operator, or a study that used historical process variation. Have the supplier identify which explanation applies and state how the next report will record it. Put that answer in the acceptance file beside the measurement method.

On an English supplier page, specify which characteristic and measurement arrangement a claimed inspection capability covers. Use a qualification example only when it is clearly scoped to the relevant line and evidence. Xindar’s manufacturing service description gives context for publishing technical evidence as readable English buyer information. Acceptance belongs with the applicable drawing, procedure and parties.

Frequently asked questions

Is percent study variation expected to add up like variance contribution?

They use different numerators. The contribution calculation uses variances. Relative study variation uses the square root of variance and the applicable study-spread convention.

Why can the percentage look better with a wider spread of parts?

The relative denominator increases. In the fictional example the measurement component remained fixed. Check the actual measurement standard deviation and percent tolerance alongside the ratio.

Can an R&R report stand in for the calibration certificate?

They provide different evidence. Use the relevant calibration record and study output to answer instrument accuracy and production measurement variation questions separately.

Which percentage belongs in the purchase order?

Agree on the decision, formula, method and limits first. State the denominator in the order so the report can be interpreted consistently by both parties.